Why the Solinst LevelVent 5 is a Great Choice:
- Vented version of our popular Levelogger®
- Accurate (0.05%FS) water level data
- 10 year battery based on 1 reading per minute
- No desiccants to replace
- Extremely durable vented cable (up to 500ft length)
- Custom blowout fitting for easy vented cable maintenance
- Complete compatibility with Levelogger accessories and Solinst telemetry systems
Where to Use the LevelVent 5:
- Ideal for deployments to 500 ft, with up to 65 ft submergence
- Excellent for aquifer characterization: pumping tests, slug tests, and more
- Stream gauging, lake monitoring, reservoir and dam management installations
- Watershed, drainage basin and groundwater recharge monitoring
- Gauged water level sensor is suited to stormwater event and runoff monitoring where rapid readings are typically required
- Water supply and tank level measurement
- Long-term water level monitoring in wells and surface water
Specifications
Model 3250 LevelVent 5 Water Level Datalogger
The Solinst LevelVent 5 is a vented version of our popular Levelogger®. The highly accurate datalogger is designed to record water level and temperature measurements during shallow groundwater and surface water deployments.
The LevelVent 5 logger contains pressure and temperature sensors, a battery, memory for up to 150,000 data logs, and hydrophobic filter within a 22 mm x 173 mm (7/8" x 6.8") stainless steel housing with a corrosion-resistant coating.
The LevelVent uses a gauged pressure transducer; it is open to the atmosphere via a vented cable to the wellhead. The compact LevelVent wellhead is designed to fit in a Solinst 2" Well Cap Assembly, and provides easy connection for communication accessories.
The vented cable and Solinst LevelVent 5 logger are protected from moisture by permanent, built-in desiccants and hydrophobic filters in the logger and the wellhead.
The vented pressure transducer is made of Hastelloy®, making it extremely durable and accurate in a wide range of temperature and monitoring conditions. The sensor provides an accuracy of 0.05% FS, and can withstand 2 times over-pressure without permanent damage.
LevelVent 5 Applications
- Ideal for shallow deployment: up to 65 ft (20 m) submergence
- Aquifer characterization: pumping tests, slug tests, etc.
- Stream gauging, lake, reservoir and dam management
- Watershed, drainage basin and recharge monitoring
- Stormwater and runoff monitoring
- Water supply and tank level measurement
- Long-term water level monitoring in wells and surface water
Vented Datalogger Advantages
- Continuous, reliable water level data for longterm monitoring projects
- No barometric compensation required, saving data processing time
- Instant water level data for real-time aquifer test results
LevelVent 5 Programming & Communication
The LevelVent 5 is programmed and data downloaded using Solinst Levelogger Software and a PC Interface Cable. It can also be used with the Solinst Levelogger 5 App Interface and DataGrabber 5. Both accessories and PC Interface Cable conveniently connect directly to the LevelVent 5 wellhead.
Integrate the LevelVent 5 with Solinst Telemetry Systems, which use the latest wireless technologies.
LevelVent Features
- Corrosion and abrasion-resistant coating baked on using polymerization technology
- Gauged pressure sensor for highly-accurate water level measurements: 0.05% FS
- Compact wellhead makes deployment simple using a Solinst 2" Well Cap Assembly
- Communication cables and accessories connect directly to the wellhead
- Long-lasting battery included in logger
- Built-in hydrophobic filters and desiccants — no need to replace, reduces maintenance
LevelVent 5 Logger Specifications
Level Sensor: | Piezoresistive Silicon with Hastelloy Sensor |
---|---|
Stability of Readings: | Superior, low noise |
Accuracy: | ± 0.05% FS Typical |
Resolution: | 0.001% FS to 0.0006% FS |
Units of Measure: | m, cm, ft., psi, kPa, bar, ºC, ºF |
Normalization: | Automatic Temperature Compensation |
Temp. Comp. Range: | 0ºC to 50ºC |
Temperature Sensor
Temperature Sensor: | Platinum Resistance Temperature Detector (RTD) |
---|---|
Temp. Sensor Accuracy: | ± 0.05ºC |
Temp. Sensor Resolution: | 0.003°C |
Battery Life: | 10 Years — based on 1 reading/minute |
Clock Accuracy: | ± 1 minute/year (- 20ºC to 80ºC) |
Operating Temperature: | -20ºC to 80ºC |
Maximum # Readings: | 150,000 sets of readings |
Memory Mode: | Slate and Continuous |
Communication: | 9600 bps Solinst USB |
Size: | 22 mm x 173 mm (7/8" x 6.8") |
Weight: | 190 grams (6.7 oz) |
Wetted Materials: | Delrin®, Viton®, Polyurethane (TPU boot) 316L stainless steel, Hastelloy, PFAS-free PTFE coating |
Sampling Modes: | Linear, Event & User-Selectable with Repeat Mode, Future Start, Future Stop, Real-Time View |
Measurement Rates: | 1/8 sec to 99 hrs |
Barometric Compensation: | Automatic |
Models | Full Scale (FS) | Accuracy | Resolution |
---|---|---|---|
M5 | 5 m (16.4 ft.) | ± 0.3 cm (0.010 ft) | 0.001% FS |
M10 | 10 m (32.8 ft.) | ± 0.5 cm (0.016 ft) | 0.0006% FS |
M20 | 20 m (65.6 ft) | ± 1 cm (0.032 ft) | 0.0006% FS |
LevelVent 5 Wellhead | |
---|---|
Materials | Black Delrin, 316 Stainless Steel, Viton |
Dimensions: | 30 mm x 76 mm (1.2" x 3.0") |
Operating Temperature: | -20ºC to 80ºC |
IP Rating: | IP 64 (dust and splash proof) |
LevelVent 5 Vented Cable Options
LevelVent 5 vented cable assemblies are available in custom lengths of up to 500 feet.
Vented cables contain communication wires, as well as a vent tube running the length of the cable. Vented cables provide direct read communication from the LevelVent 5 logger to the wellhead. The vent tube and wires are jacketed in polyurethane, providing durability and protection.
The vented cables are easily and securely connected to the LevelVent 5 logger and wellhead, using twist connections. A 2" Well Cap Assembly and hanger bracket are included to support the wellhead and vented cable when deployed.
LevelVent 5 Vented Cable | |
---|---|
Wetted Materials: | Polyurethane, Nickel plated Brass, Viton |
Diameter: | Cable: 8 mm (0.32"), Connectors: 20 mm (0.79") |
Lengths: | 1 to 500 ft |
Max. Bend Radius: | 25 mm (1") |
Operating Temperature: | -20ºC to 80ºC |
Contact Us